Dark current calibration summary 2021-01-13

Date (mean): 2021:013:18:35:16
T_ccd (mean): -8.97 C

CCD readouts summary

The table below provides the number of pixels which have the specified number of readouts. This quantifies how many pixels have been excluded either by star masking or because an entire quadrant readout is missing due to lost telemetry.

readoutsn_pixels
6269
78
829644
9520
101018135

The table below summarizes each of the 1024x1024 CCD readouts that were used to create the calibration image. Note that each readout is stitched together from four quadrant readouts that are not synchronous in time. This is reflected in the quad dt column which shows the min and max delta time from the mean.

datequad dt (sec)t_ccdint_time(s)n_masked
2021:006:20:00:08-96 to +103-10.985.0637, 698, 861, 563
2021:006:20:00:43-112 to +108-10.9810.0633, 699, 858, 563
2021:009:15:01:08-97 to +101-7.015.01135, 1294, 1128, 1391
2021:009:15:01:43-112 to +107-7.0110.01137, 1294, 1130, 1391
2021:014:18:27:26-97 to +103-9.315.01212, 778, 754, 738
2021:014:18:28:03-106 to +105-9.3110.01212, 779, 752, 739
2021:017:13:18:50-97 to +103-8.815.02245, 1609, 2182, 1775
2021:017:13:19:28-105 to +105-8.8110.02240, 1606, 2188, 1765
2021:020:02:07:17-96 to +102-8.745.02986, 2957, 2573, 2953
2021:020:02:07:52-112 to +107-8.7410.02975, 2949, 2571, 2958

Plots (all scaled to -10 C)






Reference information

See the legacy dark current calibration pages for useful information.