Flickering pixels in the ACA CCD

During analysis of the full-frame dark current calibrations, it has been observed that some warm pixels appear to "flicker" between different dark current values. Typically, a pixel would show one value for several readouts, followed by a very different value for the remaining readouts, where the change is inconsistent with counting or read noise. This behavior was seen both during the OAC (1999-Aug-11) dark calibration and the 2000-Nov-20 calibration. With dark calibration data it is possible to characterize the flickering of CCD pixels in a statistical sense. However, because of the small number of readouts (14 in OAC, and 5 in 2000-Nov), it is not clear what is happening for a particular pixel.

Fortunately, ACA data exist which contain a long time-series of data for a small number of pixels. These data come from observations with a monitor window in which the commanded location did not contain a star. Obsids 886 and 645 meet these requirements.

Obsid 886 is a long LETG observation in which the monitor window was commanded to track another guide star window. About 1/4 of the way through the observation the monitor window stopped moving, perhaps because the star in the designated track slot was momentarily lost. In any case, this is fortunate because there was a single warm pixel in the 8x8 window which was read out about 15000 times, giving a very good time history.

Dark current versus time

The plots below show the dark current versus time for 8 warm pixels found in obsid 886. Only the first one was visible during the long stationary period. The remaining seven are sampled sporadically as the dithering monitor window goes over the pixel. One can easily see the flickering nature of dark current in these pixels, wherein the dark current is essentially constant at a particular value for a period of time, and then does a discrete jump to a new value.





Dark current histograms for warm pixels

The plots below now show histograms of the dark current for each of the eight warm pixels. In some cases there appears to be a bimodal distribution, while in others (in particular the first warm pixel) there are many discrete values represented. Importantly, about 1/2 of the pixels show significant (> ~ 100 e-/sec) flickering.



Dark current histograms for normal pixels

The last set of plots below show similar histograms of dark current, but now for all pixels in a 4x6 region on the CCD. None of these pixels would be classified as warm. Many of these pixels also show distributions that are inconsistent with a single dark current value.



Distribution of flicker amplitude in dark current data
For the 2000-Nov-20 calibration, a total of 5 full-frame readouts were performed. We can estimate the number of flickering pixels by calculating the difference, for each pixel, of the median dark current value minus the minimum dark current value. For flickering pixels this value will be significantly larger than the read + counting noise. The plot below shows the distribution of this difference (the "flicker amplitude") for all 10^6 CCD pixels. There are about 600 pixels with a flicker amplitude of > 100 e-/sec. Note, however that this underestimates the true number by at least a factor of two. If most of the readouts are at the "low" value, the median will be near the "low" value, and hence the flicker amplitude will be apparently low. The problem is that median filtering is needed to screen out star images in data. Ideally, we would examine the maximum - minumum dark value. Because of possible contamination by stars, this is not feasible.

Discussion
The amplitude of flickering seen in the warm pixels is often more than 100 e-/sec. This is enough to cause noticable centroiding errors in faint stars, although these errors would not substantially degrade the final aspect solution. Nevertheless, the issue of flickering pixels needs further study. Recommendations for ACA engineering tests (with no impact to sciece) will be forthcoming.
Tom Aldcroft
Last modified: 12/15/00