ACIS SOT Shift Report See http://cxc.harvard.edu/mta/SOT_reports/sot-shift.html Report M7615. Start at 2020:148:13:10 GMT = 09:10 am EDT May 27, 2020 End at 2018:148:14:10 GMT = 10:10 am EDT May 27, 2020 ACIS SOT Shift Report May 27, 2020 ACIS Personnel: Jack Steiner, Dick Edgar, Gregg Germain, Catherine Grant, Paul Plucinsky, John ZuHone PLAN ---- We ran CAP1528 to execute a 29 hour ECS measurement during the return to science from the Safe Mode on DOY 145. The ObsID for this measurement is 62654. This is a 4 CCD ECS measurement clocking I1-I3 and S3 since a 5 or 6 CCD measurement would have violated the 1DPAMZT limit. We expect this observation will be interrupted to resume science. CLD File executed: 1A_4_CTI_39H_135.cld SUMMARY: -------- The CAP executed exactly as expected. TIMELINE (GMT day 148) ---------------------- 148:13:12 AOS confirmed on DSS-24 (connection established early) 148:13:13 Telemetry in Format 2 148:13:43 Verified radiation levels acceptable. 148:13:44 Obsid set to 62655 148:13:45 1A_4_CTI_39H_135.cld uploaded 148:13:46 SCS-135 enabled and activated 148:13:48 LOAD_TE observed 148:13:49 System configuration and Huffman table dumped 148:13:50 1STAT1ST bit changed from 1 to 0 indicating science run started 148:13:51 FP temp = -114.4 C, climbing slowly 148:13:51 1DPICACU/1DPICBCU currents at correct values 148:13:52 1DPAMZT (DPA box temp) = +20 C 148:14:02 First Bias map starts to be telemetered 148:14:08 First event data; event rates on I0-I3, & S3 look reasonable